1. the oxygen content is greater than 200ppm

Reason: Misuses.

Method: Stop cycle, open cleaning, clean it below 200ppm, turn on the loop.

2, purified column main valve fault

Cause: The gas source is insufficient.

Method: ventilation bottle, pressure is set to 0.4-0.6MPa.

3, the screen displays “pump switch off”

Cause: The vacuum pump is too large, causing trip.

Method: After repairing the vacuum pump, turn on the electric cabinet and close the trip switch.

4, cycles and analyzers turn off at the same time

Cause: The pressure in the glove box is too high, more than + 12MBAR, and the system is automatically protected.

Method: The glove will enter the glove box.

5, loop open but the fan is not turned

Cause: The voltage is unstable causes the fan inverter protection.

Method: After the voltage is stable, turn off the machine after 1 minute.

6, Main and mini antechamber can’t breathe

Cause: 1 The pumping hole in the large and small bun is blocked by a plastic film. 2 The antechamber is leaked or is not tight, or the size of the size is seized.

Method: 1 Determine the position of the pore position according to the pipeline direction, and the pipeline is cleared. 2 Tour the antechamber and check the size of the bunk seal.

Note: If the big antechamber can’t evacuate, it may be insufficient working gas pressure, and the pressure is required to be 0.4-0.6MPa.

7, cleaning function is abnormal

Cause: The pipeline behind the cleaning valve is not smooth or the working gas pressure is not open.

Method: 1 guarantee that the pipeline is smooth, don’t be bent; 2 The pressure of working gas is insufficient, and the pressure should be 0.4-0.6MPa.

8, big and small antechambercan’t make up

Cause: 1 The pores leading to the large bin in the glove box may be blocked by plastic films. 2 The pressure of the box is low, causing alarm.

Method: 1 Determine the position of the pore or dredging the piping according to the pipeline direction; 2 the box pressure is set to a positive value.