Main and mini antechambers can’t refill gas.
Cause: 1 The air hole to the main and mini antechamber in the glove box may be blocked by a plastic film;
2 Whether the solenoid valve is not opened.
Method: 1 According to the pipeline, determine the position, dredging the wire;
2 check the wire and solenoid valve.
When the main and mini antechamber can’t refill gas.
Cause: 1 The pump holes in the main and mini antechambers are blocked or the chamber is leaked by a plastic film.
2 The antechamber is not tight, or the size of the mini antechamber is loose.
Method: 1 According to the pipeline, determine the position, dredging the wire;
2 Close the antechamber door and check the size of the seal;
3 Check if the working pressure is too low (because the main antechamber pumping valve is a pneumatic valve).